Skip to content | Change text size

Monash University > RehabTech > TechGuide

Independent Information On Quality Lower Limb Prosthetic Components

Polycentric Knee Joints

Ossur (ISO 9001 Certified)

1900

Total Knee 1900 -Polymer Friction

Click for Suppliers
Related Components
No related components
Related Web Pages

No related web pages
Component Picture


A six bar friction knee. The polycentric linkage is designed to create a locking moment to prevent instability in full extension. Available with various extension bumpers. Same knee as the Model 2000 without the hydraulic control.
View Feedback on this Component

Submit Feedback on this Component
Policy on Component Feedback

Knee Specific details
Maximun Flexion Angle: 160 degrees
Knee Locking Mechanism: No
Stance Phase Control: Geometric locking system, polycentric design and adjustable stance yeilding system.
Swing Phase Control: Elastomer friction extension promoter, optional spring extension assist.

Details Established by REHABTechManufacturers Supplied Details
Compatibility:Total-knee thread to 30 mm pylon adapter Torque settings: Distal tube clamp bolt 15 Nm
Weight: 624 g Restrictions: 100 Kg
Height: 160 mm Activity Guide: No Recommendations
Material: AluminiumAssembly/ Alignment Guide: Comes with Component
Tested Limits: 100 KgWarranty: Not Available
Test standard: ISO 10328Recommended Inspection Intervals: No Recommendations
Tested By: MDA, Kungl Tekniska Hogskolan & In houseU.S.A. L-Code:
Notes: Same polycentric mechanism as the Model 2000 hydraulic knee. Also Know as Model 1900.

.
REHABTech - Centre for Biomedical Engineering
Dept. of Electrical And Computer Systems Engineering
Monash University

E-mail: rehab.tech@eng.monash.edu.au
Internet: http://www.monash.edu.au/rehabtech

Copyright © 2007 Monash University ABN 12 377 614 012 - Caution - Privacy - CRICOS Provider Number: 00008C
Last updated: - Maintained by webmaster-rehabtech@eng.monash.edu.au - Accessibility information