| Related Components 647H50, 646V43, Related Web Pages Research:  Performance of Hydraulic Knee Controllers Research:  Tester for the Decay of Prosthetic Knee Control Units Research:  Knee Characterization Data Technical:  Manufacturer's Instruction Manual Research:  Prof C.W. Radcliffe - Biomechanics of Knee Stability Control with Four-Bar Prosthetic Knees |
|
Submit Feedback on this Component
Policy on Component Feedback
| Knee Specific details |
|---|
|
Maximun Flexion Angle: 135 degrees Knee Locking Mechanism: No Stance Phase Control: Alignment Swing Phase Control: Pneumatic, flexion and extension can be adjusted independently. |
| Details Established by REHABTech | Manufacturers Supplied Details |
|---|---|
| Compatibility:Male pyramid to 30 mm pylon adapter | Torque settings: M8 15 Nm, Pylon Clamp 10 Nm, cap screws 4 Nm. |
| Weight: 816 g Knee, 186 pylon with adaptor. | Restrictions: 100 Kg Yellow |
| Height: 260 mm min, + 275 pylon and adaptor (46 ) | Activity Guide:
|
| Material: Aluminium | Assembly/ Alignment Guide: Comes with Component |
| Tested Limits: 100 Kg | Warranty: 12 Months Unconditional, 24 Months Conditional a; Only Otto Bock Components, b; Manufacturing Instructions Followed |
| Test standard: ISO 10328 | Recommended Inspection Intervals: 6 Months |
| Tested By: In House | U.S.A. L-Code: L5840 |
| Notes: discontinued | |
|
.
|