Component 3R70 no longer available and has been superceded.

Polycentric - Cadence Responsive Knee Joints

Otto Bock (ISO 9001 Certified)

3R70

Modular Polycentric Knee Joint with Pneumatic Swing Phase- Aluminium.

Click for Suppliers
Related Components
647H50, 646V43,
Related Web Pages
Research:  Performance of Hydraulic Knee Controllers
Research:  Tester for the Decay of Prosthetic Knee Control Units
Research:  Knee Characterization Data
Technical:  Manufacturer's Instruction Manual
Research:  Prof C.W. Radcliffe - Biomechanics of Knee Stability Control with Four-Bar Prosthetic Knees
Component Picture


A four bar knee with pneumatic swing phase control. The pneumatic unit is incorporated in the knee. Flexion and extension may be adjusted separately. Includes an internal adjustable extension assist. Axis ( linkage) positions may be adjusted by the prosthetist to alter the stability of the knee.
View Feedback on this Component

Submit Feedback on this Component
Policy on Component Feedback

Knee Specific details
Maximun Flexion Angle: 135 degrees
Knee Locking Mechanism: No
Stance Phase Control: Alignment
Swing Phase Control: Pneumatic, flexion and extension can be adjusted independently.

Details Established by REHABTechManufacturers Supplied Details
Compatibility:Male pyramid to 30 mm pylon adapter Torque settings: M8 15 Nm, Pylon Clamp 10 Nm, cap screws 4 Nm.
Weight: 816 g Knee, 186 pylon with adaptor. Restrictions: 100 Kg Yellow
Height: 260 mm min, + 275 pylon and adaptor (46 ) Activity Guide: Otto Bock Matrix
Material: AluminiumAssembly/ Alignment Guide: Comes with Component
Tested Limits: 100 KgWarranty: 12 Months Unconditional, 24 Months Conditional a; Only Otto Bock Components, b; Manufacturing Instructions Followed
Test standard: ISO 10328Recommended Inspection Intervals: 6 Months
Tested By: In HouseU.S.A. L-Code: L5840
Notes: discontinued

.
REHABTech - Centre for Biomedical Engineering
Dept. of Electrical And Computer Systems Engineering
Monash University

E-mail: rehab.tech@eng.monash.edu.au
Internet: http://www.monash.edu.au/rehabtech

Copyright © 2007 Monash University ABN 12 377 614 012 - Caution - Privacy - CRICOS Provider Number: 00008C
Last updated: - Maintained by webmaster-rehabtech@eng.monash.edu.au - Accessibility information